How To Select The Cuvette Material For UV

Next, it is investigated how different factors influence the selection of UV cuvette for the spectroscopic system. For example, a submicro QG quartz cell, 100uL, with a Z-size of 15 mm is requested. The window is only 5 mm high and when used for the 8.5 mm Z-dimensional instrument, the laser light source cannot pass through the window and has a light transmission of 0%. Light paths smaller than 1 mm are available, but are for more specialized buckets, such as detachable cells or HPLC flow cells.

For the purpose of photometric analyses of liquid samples, the solution must be placed in the light path of a photometer in a certain size. Buckets, i.e. sample containers with optical windows, are the default choice for this application. The distance between optical windows is precisely defined; in this way, the length of the sample path in the bucket is known. The selection of different types of cuvettes is huge, even if only those buckets used for absorption measurements in the field of UV-Vis spectroscopy are considered.

The required volume can be reduced by thickening two or four sides of the bucket inside the walls. Factors such as sample characteristics, volume availability, concentration levels, and types of measurements to be taken should influence the decision when choosing the right bin for your applications. There are so many types of buckets available, even when the range is reduced to just those used for absorption measurements in the field of UV fish spectrophotometry.

Quartz is more expensive than glass and UV transparent plastic, but it has the advantage of covering UV and visible wavelengths from 190 nm. However, it is even more fragile than glass and buckets should be handled with extreme caution. When using a UV-Vis spectrophotometer to measure absorption, choosing the right bucket is uv vis cuvettes critical, as each bucket material has its own advantages and disadvantages. The material is chosen based on the specific application of UV spectroscopy. This article provides an overview of the UV fish fry, including its material, path length, and volume. In addition, the methods in which buckets are used are discussed.

This makes them an eco-friendly option that can also save you money by eliminating the ongoing purchase of disposable plastic buckets. Excellence buckets provide exceptional window parallelism and offer small optical path length tolerances of just ± 0.01 millimeters. This enables them to deliver the precision needed for pharmaceutical and research applications. Our UV quartz submers are perfect for working within the UV-visible spectrum.

Note that there are two peaks of maximum absorption present in the dye Food Green 3, a smaller peak absorption peak at 435 nm and a more intense peak absorption peak at 619 nm. To obtain maximum sensitivity when calculating an unknown concentration of Food Green 3, the maximum absorption peak at 619 nm was used for analysis. A calibration curve was created using a linear regression equation of the least squares. The closer the data points are to a straight line, the better the setting. The intersection and in the straight line equation was set to zero to indicate that there was no absorption when no dye was present.

In the event that the sample requires high-temperature incubation, care should be taken to avoid temperatures too high for the bucket. The selection of the equipment requires requirements in the tray, as it must be compatible with the device. This mainly relates to the external dimensions of the bucket, because it must fit into the shaft of the bucket, but also the height of the measuring windows is crucial. These must perfectly match the path of light traveling through the instrument. This consideration is particularly relevant for buckets that are designed to measure small volumes and therefore have very small measuring windows. In these cases, when absorption is high and the dilution of the samples is difficult, “buckets of short path length” are applied for use.